We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Lead shielding in contact with the patient's skin is often encountered in radiotherapy with electron beams. The influence of the lead shielding on dose distributions in the patient cannot fully be assessed using modern treatment planning systems. In this work the problem of quantifying the effect of lead shielding on dose distributions is addressed. Monte Carlo dose calculations were performed in a half-blocked water phantom shielded by lead, using a realistic model for the fluence of an electron linear accelerator. Electron beam energies of 6-20 MeV and lead thicknesses of 1-7 mm are used for 10 x 10 cm2 and 5 x 5 cm2 fields. The perturbation of the particle fluence and dose distributions in water introduced by the lead shielding is quantified. The effect of oblique electron beams on the dose perturbation is shown. A fictitious clinical example, the shielding of an eye in electron beam treatment, is used to demonstrate the usefulness of Monte Carlo based treatment planning algorithms that can incorporate the effects of lead shielding.

Type

Journal article

Journal

Phys Med Biol

Publication Date

03/2001

Volume

46

Pages

757 - 769

Keywords

Algorithms, Electrons, Humans, Lead, Monte Carlo Method, Phantoms, Imaging, Photons, Radiation Protection, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Conformal, Water