We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Purpose.In small megavoltage photon fields, the accuracies of an unmodified PTW 60017-type diode dosimeter and six diodes modified by adding airgaps of thickness 0.6-1.6 mm and diameter 3.6 mm have been comprehensively characterized experimentally and computationally. The optimally thick airgap for density compensation was determined, and detectors were micro-CT imaged to investigate differences between experimentally measured radiation responses and those predicted computationally.Methods.Detectors were tested on- and off-axis, at 5 and 15 cm depths in 6 and 15 MV fields ≥0.5x0.5 cm2. Computational studies were carried out using the EGSnrc/BEAMnrc Monte Carlo radiation transport code. Experimentally, radiation was delivered using a Varian TrueBeam linac and doses absorbed by water were measured using Gafchromic EBT3 film and ionization chambers, and compared with diode readings. Detector response was characterized via thekQ clin ,Qmsrfclin,fmsrformalism, choosing a 4x4 cm2reference field.Results.For the unmodified 60017 diode, the maximum error in small field doses obtained from diode readings uncorrected bykQ clin ,Qmsrfclin,fmsrfactors was determined as 11.9% computationally at +0.25 mm off-axis and 5 cm depth in a 15 MV 0.5x0.5 cm2field, and 11.7% experimentally at -0.30 mm off-axis and 5 cm depth in the same field. A detector modified to include a 1.6 mm thick airgap performed best, with maximum computationally and experimentally determined errors of 2.2% and 4.1%. The 1.6 mm airgap deepened the modified dosimeter's effective point of measurement by 0.5 mm. For some detectors significant differences existed between responses in small fields determined computationally and experimentally, micro-CT imaging indicating that these differences were due to within-tolerance variations in the thickness of an epoxy resin layer.Conclusions.The dosimetric performance of a 60017 diode detector was comprehensively improved throughout 6 and 15 MV small photon fields via density compensation. For this approach to work well with good detector-to-detector reproducibility, tolerances on dense component dimensions should be reduced to limit associated variations of response in small fields, or these components should be modified to have more water-like densities.

Original publication

DOI

10.1088/1361-6560/ab91d9

Type

Journal article

Journal

Phys Med Biol

Publication Date

11/05/2020

Keywords

density compensation, detector response, diode, dosimetry, small field